Figure 44 Eo Box - Clarity RetCam3 User Manual

Ophthalmic imaging systems
Table of Contents

Advertisement

70
Chapter 4
2. Turn the main power switch, located on the front of the EO box, to ON (unless already
ON).

Figure 44 EO Box

3. Plug the main power cord from the system into a properly grounded hospital grade
electrical outlet. Turn ON main power at the back of the RetCam Shuttle cart. The
green power lamp indicator illuminates and you hear the internal fan. Power ON the
notebook computer. The system is ready when you see the User Login screen
(Figure 46 on page
RetCam 3 and RetCam Shuttle User Manual
72).
Connect this equipment using the power cord supplied by Clarity or agency
listed power cord for the country of use.
To avoid the risk of electric shock, this equipment must be connected to a
supply main with protective earth.
NOTE
The system has no internal power supply (no battery) to support image
acquisition if disconnected from a wall outlet. (The notebook computer can
operate using its own battery.)
Preparing the System for Use
EO Box power switch
PN 18-000393 Rev. A