136
Chapter 10
Figure 134 New Exam Configuration Screen
2. To capture a color image (recommended), click
label has changed to
3. Switch ON FA power. The FA lamp requires a few minutes to stabilize
page
4. Click
image (recommended).
5. Carefully install the barrier filter into the handpiece
a. Remove the lens piece (if present).
b. Insert the barrier filter into the cavity of the handpiece, observing the orientation as
c. Reattach the lens piece before moving the handpiece.
RetCam 3 and RetCam Shuttle User Manual
138).
. Confirm the button label has changed to
NOTE
Certain FA models include a warning LED between the Fiber Optic connector
and the intensity adjustment knob.
shown in
Figure 135 on page
Fluorescein Angiography (RetCam 3 System)
.
(Figure 135 on page
137.
and confirm that the button
(Figure 136 on
. Capture a "red-free"
137):
PN 18-000393 Rev. A