Table 1 Retcam 3 Electromagnetic Emissions - Clarity RetCam3 User Manual

Ophthalmic imaging systems
Table of Contents

Advertisement

26
Chapter 1

Table 1 RetCam 3 Electromagnetic Emissions

Test Type
Radiated Emissions
CISPR 11/
EN 55011
Conducted Emissions
CISPR 11/
EN 55011
Harmonic emissions
EN 61000-3-2
Voltage fluctuations/flicker emissions
EN 61000-3-3
FCC Part 15B
RetCam 3 and RetCam Shuttle User Manual
Compliance Level: Group 1, Class A
Compliance Level
Group 1
Class A
Class A
Complies
Complies
Notes
The RetCam system uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment. The RetCam system is suitable for use in all
establishments other than domestic, and may be used in
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes, provided the
following warning is heeded:
Warning: This equipment/system is intended for use by
health care professionals only. This equipment may cause
radio interference or may disrupt the operation of nearby
equipment. It may be necessary to take mitigation
measures, such as re-orienting or relocating the system
or shielding the location.