7 Maintenance and Support

This section addresses maintenance and support under the following topics:
Recommended Maintenance
Cleaning
Procedures, below

Recommended Maintenance Schedule

Between patients: Clean and inspect lens piece.
Weekly: Wipe down the system.
Monthly: Inspect cables and connections for wear.
System powered for at least 8 hours (AC powered - AC/DC indicator ON
light solid) to assure backup battery is fully charged.
Burned out bulb: Replace illumination bulb.
Blown fuse: Replace system fuses.

Cleaning Procedures

For cleaning purposes, the RetCam3 is divided into two categories:
• The Lens Piece that contacts the patient
• The rest of the system

Cleaning the Lens Piece

The patient contact area lens should be cleaned immediately after use to prevent
the coupling gel from hardening.
1. Wipe away debris from the front lens piece surface with a soft dry tissue, sterile
water soaked tissue, or combination.
2. Then, using a clean soft cloth (such as sterile gauze) saturated in a fresh
solution of 70% IPA (isopropyl alcohol), gently wipe the front of the lens piece,
being sure to pay special attention to the concave contact lens area, to remove
any remaining debris. Use at least 11 wipes with a clean saturated cloth.
3. Rinse surface thoroughly using sterile water.
4. Air dry.
Schedule, below
59
61
65
66
67
PN 21-100375 Rev. A