• Prior to the start of measurement, be sure to perform white calibration. For details, refer to "White
Notes
Calibration" on page 34.
• To display the color difference, target colors must be set before measurement.
• To measure a target, select the target number before measurement.
• For accurate measurement, make sure to keep measurement conditions (ambient temperature, etc.)
constant.
Operating Procedure
1
Press [
] to bring up the <Sample> screen.
The <Sample> screen is displayed.
If the menu screen is displayed, press [ESC] and perform the
Memo
operation after the Results Display screen is displayed.
2
Set the specimen measuring port on the sample to
measure.
Measure with the cross on the base of the specimen
Memo
measuring port on the sample.
3
Press the measurement key or the measurement
button.
The specimen is measured and the results are displayed on
the screen.
• The measurement data No. is automatically assigned
sequentially in the order of measurement.
• When the number of measurement data sets reaches 800, an error message will be displayed and measurement will
Memo
not be possible. In such cases, delete some data to perform the measurement.
• If the screen is not displayed because the instrument is in sleep mode, measurement will not be possible. Press any
of the control keys to display the screen, and then press the measurement key/button to perform measurement.
Sample
0002 TARGET-2
0059
No Name
0.24
10°/D65
∆FF
1
2
2°/C
L*
C*
h
97.42
0.30 224.43
-15°
98.70
0.16 232.93
15°
98.63
0.21 259.31
25°
98.68
0.06 140.84
45°
98.29
0.02 118.77
75°
94.67
0.13 106.71
110°
2016/08/23 16:28:14
0001 TEST-1
MENU
ESC
Sample
0002 TARGET-2
0060
No Name
0.24
1
10°/D65
2
2°/C
∆FF
L*
C*
h
97.41
0.31 224.85
-15°
98.70
0.16 234.00
15°
98.62
0.23 261.20
25°
98.67
0.05 144.71
45°
98.28
0.02 106.89
75°
94.67
0.14 103.00
110°
2016/08/23 16:28:14
0001 TEST-1
2
36

Advertisement

Table of Contents
loading

Table of Contents