Pass/Fail Judgment for Color Difference

With this instrument, tolerances can be set for the color difference of the sample data from the target color data to
perform judgment. For the procedure on setting tolerances, refer to page 55 "Tolerance Settings" and page
81 "Default Tolerance Setting".
The pass/fail judgment will be based on the target data of the number selected for measurement and on the
tolerance data specified for that target data. If target color data is deleted, neither the color difference display for
the data nor the pass/fail judgment based on the tolerances set to the data will be performed. If other target color
data is selected, or another target color is assigned to the target color number of the deleted data, recalculations
and pass/fail judgments will be performed.

Pass/Fail Judgment Based on Tolerances

If the measured color difference is outside the tolerances set for the target color, the value will be highlighted in
red to indicate that the judgment result is "Fail". Upper and lower tolerances can be set for color differences of
individual targets.
The following section details how to display the pass/fail judgment based on the color difference tolerance.
Before using this function, the color difference tolerances must be set.
Notes
To measure without using a target, set a number that has not
Memo
been registered with a target as the target. This will provide
absolute values, so color difference calculation and pass/fail
judgment per the tolerance will not be possible. The Pass/
Fail screen will be displayed as shown at right.
39
Sample
0002 TARGET-2
0002
No Name
10°/D65
2°/C
1
2
CMC
∆L*
∆C*
∆H*
Warning
2016/08/23 16:28:14
0001 TEST-1