White Calibration

This instrument requires white calibration before performing measurement after the power is turned ON and
before performing measurement for the first time.
• If the calibration interval is turned ON and a time is configured, a message prompting white calibration will be
Memo
displayed the next time the power is turned ON only after the set time has passed since the previous white
calibration. (Refer to page 85 "Calibration Interval Messages".)
• The reading may fluctuate slightly due to changes in the ambient temperature or due to heat generation caused by
repeated operation of the instrument. In such cases, perform white calibration regularly.
• White calibration must be performed at the same temperature at which measurement will be performed.
Notes
• Perform white calibration after the instrument has had time to adapt to the ambient temperature.
Operating Procedure
Start the procedure from the measurement screen.
Although white calibration can be performed from the prompt when the power is turned ON and from the screen
following zero calibration, the following explains the procedure to perform white calibration from the
measurement screen.
1
Press [MENU], and then use [] or [] to display the
<Calibration> screen.
• The following screen prompting calibration is
Notes
displayed when the instrument starts up.
When power is turned ON
2
Use [] or [] to select "White calibration", and
then press the [Confirmation] key.
3
Attach the white calibration cap to the instrument
while referring to page 20 "White Calibration
Cap".
Calibration

White calibration

Calibration
(Including Zero Cal.)
Zero calibration date
2016/08/22 13:51:58
White calibration date
2016/08/23 09:30:10
White ID
1000001
Sample
Calibration
White calibration
Calibration
(Including Zero Cal.)
Zero calibration date
2016/08/22 13:51:58
White calibration date
2016/08/23 09:30:10
White ID
1000001
Calibration
Please set the
instrument on the white
calibration plate and
press the measurement
button.
White ID 1000001