Measurement Procedure

c This manual explains how to prepare the CF-300 and how to set a specimen.

c The CF-300 is controlled by a PC to perform measurements.

c For a description of measuring method using SpectraMagic DX (optional),

refer to the SpectraMagic DX instruction manual.

 Flow of Preparation and Measurement


Connecting the PC
Connecting the AC adapter
Starting the PC
(Starting up Windows ®)
Starting the software
Turning the power ON
Attaching a target mask
Performing zero
calibration
Performing white
calibration
Positioning a specimen
Performing measurement
Turning the power OFF
Connect the CF-300 to the PC with the USB cable. (P.13)
Connect the CF-300 to the AC outlet
with the AC adapter. (P.14)
Turn on the PC to be used to control the CF-300.
Start the software and set it for control of the CF-300.
Turn the power ON. (P.15)
Attach the target mask (CM-A91) for calibration. (P.16)
Position the zero calibration box and
perform zero calibration. (P.18)
Position the white calibration plate and
perform white calibration. (P.19)
Position the specimen on / in the CF-300. (P.20)
Perform measurements.
When measurements are complete, turn the power OFF.(P.15)
Exit the software and turn off the PC.
12