Measurement Procedure

c This manual explains how to prepare the CM-3700A-U and how to set a specimen.

c The CM-3700A-U is controlled by a PC to perform measurements.

c For a description of measuring method using SpectraMagic™ NX

(optional), refer to the SpectraMagic™ NX instruction manual.

 Flow of Preparation and Measurement


Connecting the PC
Connecting the AC adapter
Starting the PC
(starting Windows)
Starting the software
Turning the power ON
Attaching a calibration
target mask
Performing zero calibration
Performing white calibration
Attaching a measurement
target mask
Positioning a specimen
Performing measurement
Turning the power OFF
Connect the CM-3700A-U to the PC
with the USB cable. (P.13)
Connect the CM-3700A-U to the AC outlet
with the AC adapter. (P.14)
Turn on the PC to be used to control the CM-3700A-U.
Start the software and set it for control of the CM-3700A-U.
Turn the power ON. (P.15)
Attach the target mask (CM-A91) for calibration. (P.16)
Position the zero calibration box and
perform zero calibration. (P.19)
Position the white calibration plate and
perform white calibration. (P.20)
Attaching an appropriate measurement target mask.
Position the specimen on / in the CM-3700A-U. (P.21)
Perform measurements.
When measurements are complete, turn the power OFF. (P.15)