Pass/Fail

Edit the tolerance that will be used as the judgment criteria, and set the warning level and the parametric
coefficient default value.
If the color difference between the sample and the target exceeds the tolerance, the column of the display color
value relevant to the measurement display will be displayed in red. In addition, if even one display color value
exceeds the tolerance, judgment will be shown as "Fail".
If the color difference between the sample and the target exceeds the warning level amount, the column of the
display color value relevant to the measurement display will be displayed in yellow. In such cases, even if other
display color values do not exceed the tolerance, judgment will be shown as "Warning".
• The default judgment criteria is set before judgment criteria for each target is set. For details, refer to page 80
Memo
"Default Data Settings".
• The tolerance setting items are the same as those for the color space or index currently selected for the display
conditions.
• The optional Color Data Software "SpectraMagic DX" allows for easy configuration and use of judgment criteria.
• Display the target to be set in advance before moving to <Target menu>.
Operating Procedure
Start the procedure from the <Target menu> screen.
1
Use [] or [] to move the cursor to "Pass/Fail",
and then press the [Confirmation] key. The <Pass/
Fail> screen will be displayed.
Target
Target menu [0002]
Delete data
Edit name
TARGET-2
Data management

Pass/Fail

Target
Target menu [0002]
Pass/Fail
Edit tolerance
Warning level
80%
Parametric coef.
3