4.4
Connecting the Patient Circuit
WARNING!
Read the chapter 2.4 Usage of Patient Circuit, page 15 carefully to make sure all conditions
are considered and met.
NOTE
•
"Select Patient Circuit Type" is located in the Others section of the Vivo 65 display when
the ventilator is in Standby mode.
•
In MPV breath mode the patient circuit type setting is not available. Circuit with mouth-
piece interface shall always be used in MPV breath mode.
The Vivo 65 can be used with the following circuits:
•
Dual limb circuit (Adult use, 22 mm), connected to the Vivo 65's integrated exhalation
valve
•
Dual limb circuit (Paediatric use, 15 mm), connected to the integrated exhalation valve
•
Single limb circuit with external active exhalation valve; requires an optional single limb
insert
•
Single limb circuit with external leakage port
•
Circuit with mouthpiece interface
4.4.1
Connect Patient Circuit Insert
The choice of insert makes it possible to connect the Vivo 65 to either a dual limb paediatric
circuit, a dual limb adult circuit or a single limb circuit. The Vivo 65 automatically detects
which insert has been installed. Before connecting an insert, make sure the configurations of
patient mode and patient circuit type are correct:
34
Preparing the Vivo 65 for Use
User manual
Make sure that the correct patient circuit type is selected when connecting the
insert and patient circuit.
Doc. 006150 M-1