Change of Patient

If the Clearway 2 is used in a clinic by several patients, use of a
low resistant bacterial filter is essential between the air outlet
and the patient tube to prevent patient cross-contamination.
1.
Follow the instructions in "Cleaning the Clearway 2" (page 57).
2.
Replace the patient filters according to "Changing the Air Inlet Filter"
(page 59).
3.
If a low resistant bacterial filter is used, it shall be replaced regularly.
4.
Use a cleaned or a new patient circuit and a new dual insert when the
Clearway 2 is used by a new patient.
Cleaning a device suspected to contain pathogens cannot be
carried out by the lay operator in the homecare environment.
The device must be returned to the manufacturer for disposal or
disposed of in the correct way (see Disposal, page 62).
Clearway 2 is not suitable for sterilisation.
DOC-001000 Ver 6
Cleaning & Maintenance
Clearway 2 User Manual
60
.