Alarms - Dräger Evita 2 Instructions For Use Manual

Intensive care ventilator
Table of Contents

Advertisement

Betrieb
Alarms
1 The
red alarm light flashes, accompanied by alarm
sound.
2
Alarm message is displayed on the screen.
l
Rectify fault, use 'Fault-Cause-Remedy'
list, page
104to106.
When fault is rectified, the alarm sound ceases.
Acknowledging
alarm message:
3
Press @%l button, the alarm message disappears
from the screen.
The alarm sound can be suppressed for 2 minutes:
4
Keep @?I button pressed until the yellow LED is lit.
02 alarm
Calibrate 02 sensor, page 17,
or:
replace 02 sensor, page 8.
Switching
off 02 monitoring
in exceptional
cases
5
Press B=G button. yellow LED flashes:
lzzl
01 monitoring
is switching
off permanently.
Switching 02 monitoring on again:
5
Press .M
button again, yellow LED goes out.
rl
80