9. Multiphor II application kits and accessories
Cut a 10 cm length of silicone tubing and make a 2 cm long cut across one
end of the tubing.
Note: It is important that the cut is 2 cm long since it may be difficult to
get the tubing through the hole in the glass plate with a shorter cut.
Hold the U-framed glass plate with the spacer facing the user. Push the cut
end of the tubing through the hole in the bottom corner of the plate. Pull the
tubing through the hole until the other end is flush with the inner face of the
plate. Apply the pinch cock to the silicone tubing. If required, the tubing can
be permanently fixed using a silicone glue, such as General Electric RTV
Silicone Rubber Adhesive Sealant 108.
The following instructions apply to Immobiline DryPlate. If using
Immobiline DryStrip, see the instruction manual supplied with Immobiline
DryStrip Kit.
Clean the 3 mm thick glass plate and lightly wet it with a few drops of
water. Place the dry gel on the glass plate with the gel side up. Roll the gel
flat with the roller to remove air bubbles. Keep water from getting onto the
86