5. Operation
36
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Sample application
Use the same methods as for Ampholine PAGplate (see Section 4.4).
Running conditions
In principle, CleanGel is run in the same way as Ampholine PAGplate
except no electrode strips are used. The electrodes rest directly on the gel
surface.
Recommended running conditions for one CleanGel IEF 3-10
Phase
Voltage
(V)
Prefocusing
700
Sample
entrance
500
Isoelectric
focusing
2000
Band
sharpening
2500
Note: If only half a gel is used, halve the current and power settings.
Detection Methods
Use the detection methods described for Ampholine PAGplate
(see Section 4.4).
Current
Power
(mA)
(W)
12
8
8
8
14
14
14
18
Time
(min)
20
20
90
10