4.2 Multiphor II
NovaBlot Unit
for electro-
phoretic
transfer
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Replace the nut and lightly tighten until the electrode is held firmly in place.
To align the electrodes with the electrode strips, place the electrode holder
with the electrodes onto the false holes on the buffer tank using the four
corner feet.
Install the safety lid.
Multiphor buffer tank should be assembled with the four levelling feet.
Make sure the buffer tank is rinsed with water and dried.
Saturate the anode NovaBlot electrode with distilled water and remove the
excess water with absorbent paper. With the electrode lead to the front of the
instrument, fit the anode electrode onto the buffer tank.
Connect the anode socket (red lead) to the anode pin on the right side of the
buffer tank.
Place a spirit level on the anode electrode and adjust the levelling feet until
the unit is levelled.
4. Installation