4. Installation
For all products, check the unassembled parts against the Packing List for
the respective product to ensure that all items heve been included.
For easy connection of Multiphor II Electrophoresis Unit to the cooling
device, install tubing connectors on both inlet and outlet tubings. Use one
male and one female connector on the Multiphor II unit side and thermostat
unit side respectively. The cooling plate tubings and thermostatic circulator
tubings can then be locked separately.
Two Multiphor II units can also be connected in series to one thermostatic
circulator.
Place the rubber grommets on the cooling plate inlet and outlet tubes. Slide a
short piece of tubing onto each tube and secure with hose clamps. Attach the
male and female tubing connectors as described above. Repeat this process
with the thermostatic circulator using longer pieces of tubing.
To lock the connectors, insert the male connector into the female and turn
clockwise one-quarter turn until it clicks.
Screw one levelling foot into each corner of the buffer tank. Place the buffer
tank on the lab bench where it will be used. Place the cooling plate on the
unit, using the moulded guides to position it correctly. Fit the grommets into
the cutouts in the back of the unit. Place a spirit level on the cooling plate
and adjust the levelling feet until the unit is levelled.