Running Conditions For Precast Gels - Pharmacia Biotech Multiphor II User Manual

Electrophoresis system
Table of Contents

Advertisement

5. Operation
5.10 Running
conditions
for precast
gels
48
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Transfer buffers using a discontinuous buffer system
R. Anode solution 1, pH 10.4
Tris
36.3 g
Methanol
200 ml
Make up to 1000 ml with distilled water.
S. Anode solution 2, pH 10.4
Tris
3.03 g
Methanol
200 ml
Make up to 1000 ml with distilled water.
T. Cathode solution, pH 7.6
6-Amino-n-hexanoic acid
Methanol
Make up to 1000 ml with distilled water
U. Transfer buffer using a continuous buffer system
Glycine
2.93 g
Tris
5.81 g
SDS
0.375 g
Methanol
200 ml
Make up to 1000 ml with distilled water.
Note: In a continuous buffer system, this solution is used for both anode
and cathode electrode solutions.
ExcelGel SDS gradient 8-18
Running conditions
Phase
Voltage
Current
(V)
(mA)
Run
600
50
* Or until the Bromophenol Blue front reaches the anode buffer strip.
ExcelGel XL SDS gradient 12-14
Running conditions
Phase
Voltage
Current
(V)
(mA)
Run
1000
40
* Or until the Bromophenol Blue front reaches the anode buffer strip.
ExcelGel SDS Homogeneous 7.5, 12.5 and 15
Running conditions
ExcelGel SDS
Voltage
Homogeneous
(V)
7.5 and 12.5
600
15
600
* Or until the Bromophenol Blue front reaches the anode buffer strip.
5.20 g
200 ml
Power
Time
(W)
(min)
30
75 *
Power
Time
(W)
(min)
40
165 *
Current
Power
(mA)
(W)
50
30
30
30
Temp
°C
15
Temp
°C
15
Time
Temp
(min)
°C
80*
15
140*
15