4. Installation
18
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The transfer sandwich can now be assembled on the bottom anode electrode.
Full assembly instructions are given in Section 4.8.
After assembling the transfer sandwich, saturate the upper cathode electrode
with distilled water and remove any excess with absorbent paper. With the
electrode lead to the front of the instrument, place the cathode electrode on
top of the stack.
Connect the cathode socket (black lead) to the cathode pin on the left of the
buffer tank.
Install the safety lid.
WARNING! Check that the electrodes are properly connected
before closing the lid. Always check that the lid is properly before
turning the power supply ON.