Connecting Cooling Water - Advanced Energy Paramount MF 2 kW User Manual

Generator
Table of Contents

Advertisement

®
Advanced Energy
suitable chassis ground connection made to this stud prevents or minimizes
radio frequency interference.

Connecting Cooling Water

This generator is water cooled. Do not operate the unit until you have connected
cooling water and met the cooling requirements.
  CAUTION:
If you connect the cooling water on multiple units in series, be sure that input
water temperature to all units is less than the maximum input water
temperature.
  CAUTION:
If you connect the cooling water on multiple units in a series, be sure that the
water flow rate to all units meets the unit specifications.
  CAUTION:
Do not use deionized water for cooling purposes. Deionized water causes
both corrosion and erosion of cooling manifolds.
  CAUTION:
Do not apply more than 54.2 Nm (40 ft-lb) to water fittings. Water leakage
and damage to the unit could result.
The water fittings used on the Paramount MF 2 kW generator are SS female 3/8"
Swagelok.
TO CONNECT COOLING WATER
1. Make input and output water connections and tighten securely.
2. Turn on the water and ensure that there are no leaks.
3. Be sure that the flow rate, pressure, and temperature are within the minimum
specifications required to operate the generator.
57020114-00E
Installation, Setup, and Operation
®
Paramount
MF 2 kW Generator
5‑11