Table 3‑2. Electrical specifications (Continued)
Description
Diagnostics
Tuning Algorithm
The unit is designed and programmed to utilize a sweep frequency algorithm to optimize power
delivery and minimize load mismatch as detected by the unit's output sensor. The algorithm is
able to:
• Minimize load mismatch
• Lock the generator at a tune point
• Track (re-tune) frequency to follow changes in load impedance to minimize load mismatch
Frequency tuning range
Frequency preset: Start
minimum and maximum
frequency
Tune time
Typical tune time for
factory set tuning
parameters
Tracking and re-tune
Tuned frequency
repeatability
Pulsing into Broadband 50 Ω
Frequency range
Duty cycle range
Minimum on width
Rise/ fall time
Settling time
Voltage overshoot
57020114-00E
The generator will evaluate its internal parameters for the relative
health of the generator. Special diagnostic capability is provided for
testing into short (0 Ω) load.
360.1 kHz to 439.6 kHz
360.1 kHz to 439.6 kHz, 1 kHz resolution
< 0.1 second from RF enabled at a preset frequency (within 10% or
2 W of the minimum reflected power, whichever is greater)
60 ms tuning delay +25 ms tuning
< 0.1 s maximum
If load mismatch increases above a user defined threshold, the
generator will re-tune.
± 2 kHz
10 Hz to 2 kHz (minimum on width has to be satisfied)
10% to 90% (minimum on width has to be satisfied)
225 μs
When output voltage reaches 67% of nominal:
• Rise time 10 μs maximum
• Fall time 10 μs maximum
When output voltage settles within 10% of nominal:
• 300 μs at 5 W output power
• 30 μs at 2000 W output power
With a 50 Ω broadband load:
Specifications
®
Paramount
MF 2 kW Generator
Specification
3‑7