5 Operation
5.5 Main home screen
40
Surgery
Abscess
Epulis
Fibroma
Fistula
Frenectomy
Gingivectomy
Gingivoplastic
Haemostasis
Implant Uncovery
Incisions/Excisions
Operculectomy
Gingival Troughing
1. Select the desired indication.
Ä The pre-set parameters will be displayed.
2. Now you can activate the laser: Press the 'Activate Laser' button.
Ä You will be advised to wear protective goggles before the
aiming beam is activated.
3. Acknowledging the advice.
Ä The green LEDs start flashing. After a delay of 2 seconds, the
aiming beam is switched on.
Ä The laser is now ready for operation.
WARNING
All persons present in the room (operator, assistants and patient)
must wear the laser protective goggles as soon as it is advised to
wear laser protective goggles and whenever the green LEDs are lit.
Any actuation of the finger switch activates the laser unit.
Wrong settings may result in severe damage of the patient's soft or
hard tissue or may result in no treatment efficacy. So any user is
supposed to have sufficient knowledge and training in laser therapy.
The treatment room must be protected by suitable measures (in
compliance with IEC 60825-1), e.g. by closing the doors.
NOTE
Before starting a laser treatment in battery operation please reconfirm
the battery status.
When you activate the finger switch, the laser begins emitting. At the
same time, two yellow LEDs at the upper right and left end of the
SIROLaser Xtend control unit light up as well as the laser active bar on
the touch screen and the audible alarm sounds. When you release the
finger switch to interrupt treatment, the laser is deactivated, but remains
ready for operation.
Sirona Dental Systems GmbH
Operating Instructions SIROLaser Xtend
63 18 351 D3576
D3576.201.01.05.02    11.2018