Sirona SIROLaser Xtend Operating Instructions Manual page 37

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Operating Instructions SIROLaser Xtend
63 18 351 D3576
D3576.201.01.05.02    11.2018
NOTE
If the USB port is marked 'cross' for defective please contact your
local dental depot or an authorized Customer Service Department for
technical support.
The SIROLaser Xtend will remain functional but the download of the
history file is not possible.
Calibration check
Sirona Dental Systems recommends a calibration check with external
powermeter to be performed every twelve months, see chapter
"Calibration check"  [→ 46].
'Next calibration check' counts down from twelve to zero and will be
highlighted if zero is reached. The laser remains fully functional. Unit is
month.
Time to service
The safety test is a mandatory test for all medical devices. The
SIROLaser Xtend needs to be tested once every two years.
"Time to service" counts down from 24 to zero and will be highlighted if
zero is reached. The laser remains fully functional. Unit is month.
NOTE
Legal regulations require a regular safety test of the performance of a
laser device. The SIROLaser Xtend needs to be tested once every
two years. Please contact your local dental depot or an authorized
Customer Service Department for technical support.
If internal or external calibration check failed after last calibration a
warning screen pops up after self test. The laser remains fully
functional.
NOTE
Please contact your local dental depot or an authorized Customer
Service Department for technical support.
1. To redo the self test press the 'back' button.
2. To quit the self test press 'OK'.
5 Operation
5.3 Self Test
37