Sirona SIROLaser Xtend Operating Instructions Manual page 71

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Operating Instructions SIROLaser Xtend
Interference
IEC 60601-1-2 test level
immunity tests
Conducted RF
3 V
interference
150 kHz to 80 MHz
IEC 61000-4-6
Radiated RF
3 V/m
interference
80 MHz to 800 MHz
IEC 61000-4-3
3 V/m
800 MHz to 2.5 GHz
63 18 351 D3576
D3576.201.01.05.02    11.2018
Compliance level
3 V
eff
1
3 V
1
3 V
1
1. The higher frequency range applies at 80 MHz and 800 MHz.
2. Field strengths from fixed transmitters, such as base stations for
radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast, cannot be
predicted theoretically with accuracy. An investigation of the
location is recommended to determine the electromagnetic
environment resulting from stationary HF transmitters. If the
measured field strength in the location in which the UNIT is used
exceeds the applicable RF compliance level above, the UNIT
should be observed to verify normal operation. If unusual
performance characteristics are observed, it may be necessary to
take additional measures such as reorientation or repositioning of
the UNIT.
3. Over the frequency range 150kHz to 80MHz, field strengths should
be less than 3V/m.
Electromagnetic environment –
guidance
d= [1.2] √P
eff
d= [1.2] √P
eff
at 80 MHz to 800 MHz
d= [2.3] √P
eff
at 800 MHz to 2.5 GHz
where P is the nominal transmitter
output in watts (W) specified by the
transmitter manufacturer and d is
the recommended working clear-
ance in meters (m).
Field strengths from fixed RF trans-
mitters, as determined by an electro-
magnetic site survey
than the compliance level
frequency range.
Interference is possible in the vicinity
of equipment bearing the following
graphic symbol.
9 Electromagnetic compatibility
9.2 Interference immunity
2
should be less
3
in each