7 Cleaning, disinfection and sterilization

7.2 Disinfection

62
CAUTION
Danger of damage to the optical system
Reattach immediately after dismounting the fiber the protective caps
to the optical system of the handpiece and to the proximal end of the
ferrule of the fiber. Do this before taking any hygienic measures.
6. Clean the handpiece sleeve with a suitable brush under running
water.
Fibercutter
➢ Clean the fiber cutter in the ultrasonic bath or with a suitable brush
under running water.
Laser protective goggles
➢ Before cleaning the laser protective goggles, please read and
observe the instructions for use provided by the manufacturer and
attached to the goggles in the case.
7.2
Disinfection
Disinfect the upper mentioned parts by wipe disinfection:
SIROLaser Xtend (wipe disinfection only)
NOTE
Use only disinfectants that comply with the requirements of your
national authorities and whose bactericidal, fungicidal and virucidal
properties have been tested and properly certified.
Sirona recommends the use of MinuteWipes from Alpro. In the USA:
Caviwipes
.
Observe the instructions provided by the manufacturers of these
disinfectants.
CAUTION
The optical fiber may be damaged if it is seriously bent or improperly
routed inside the handpiece. This may constitute a health hazard for
patients, dentists and dental assistants. The minimum bending radius
of the optical fiber is 4.5 cm (diameter: 9 cm). Be careful not to pinch
or tear the optical fiber when inserting or cleaning it.
Sirona Dental Systems GmbH
Operating Instructions SIROLaser Xtend
63 18 351 D3576
D3576.201.01.05.02    11.2018