Operating Instructions SIROLaser Xtend
63 18 351 D3576
D3576.201.01.05.02    11.2018
1. Open handpiece by pressing both snap tabs and pull apart the
handpiece sleeve from the handpiece body.
2. Remove the protection screw from the handpiece body by using an
empty transport and sterilization tube.
3. Remove the protective cap from the optical fiber.
4. Insert the optical fiber into the handpiece and screw it on. Tighten
the optical fiber only slightly so that it can be loosened with the
transport and sterilization tube again with ease. Make sure not to
overwind the transport and sterilization tube!
5. Remove the transport and sterilization tube from the optical fiber.
OK
6. Put the laser into operation by choosing any preset treatment. A
corresponding description is provided in chapter "Operation" 
[→ 32].
The aiming beam must not be aimed toward a person's eye. It
comprises an intensive light source even when set to a low power
level. Always wear protective goggles.
7. As soon as the SIROLaser Xtend is ready you can check to make
sure that the aiming beam illuminates evenly, i.e. projects a circular
light pattern (see adjacent picture).To do this, aim the optical fiber
vertically at a white background located approx. 5 cm away. If the
beam shows no pattern or the beam pattern is not illuminated
evenly, the optical fiber may be damaged or defective. In this case,
return the optical fiber to your dental depot so that it can be
replaced under warranty. Do not use any defective optical fibers.
4.4 Initial start-up – procedure for proper assembly
WARNING
4 Installation
27