Operating Personnel; Physical Working Principle; Functional Check - Sirona SIROLaser Xtend Operating Instructions Manual

Table of Contents

Advertisement

2 Safety information

2.3 Operating personnel

Qualification/education
Know-how
Experience
Obligation of the user
Unauthorized access

Functional Check

8
Operating personnel
2.3
The SIROLaser Xtend may only be operated by educated and qualified
personnel (dentist, assistent, dental hygienist). The applicable
occupational safety regulations and accident prevention measures, the
current operating instructions and national requirements concerning
education must be complied with.
Know-how and expertise about laser therapy as well as the skilled use
of the laser and the applied indications are required. Please refer to
applicable country-specific requirements.
Sirona recommends to gain practical experience in laser dentistry
before first use of the SIROLaser Xtend by attending an appropriate
training. Amongst others Sirona offers trainings. Please see the Sirona
homepage (www.sirona.com).
Users are obliged to use only faultless materials, to ensure correct
application and to protect themselves, the patient and other persons
against hazards.
In order to prevent false or improper use, the SIROLaser Xtend must
not be used by unauthorized persons. Therefore the SIROLaser Xtend
equipment must be protected against unauthorized access when not in
use. This can be achieved, for example, by switching the SIROLaser
Xtend off after usage so that the electronic access key (pin code) must
be entered before using it again.
WARNING
The SIROLaser Xtend may only be used and maintained by
thoroughly trained personnel.

Physical working principle

2.4
The 970 nm laser radiation of the SIROLaser Xtend is generated via a
laser diode inside the control unit and guided to the treatment region via
quartz fibers. The laser radiation is absorbed by the tissue and
converted to heat used for cutting, coagulation, germ reduction and
desensitization.
Functional check
2.5
The system owner is obliged to use only technically faultless products.
Please check the unit for proper functioning every time before putting it
into operation. In case of unusual noises, check both the unit and the
handpiece. If the unit has fallen down, have it checked by qualified
technical personnel.
To prevent the unit being accidently pulled from the table, the
handpiece hose should never be under tension. Please always ensure
that ca. 40 cm of the handpiece hose hangs.
If there is any doubt about the correct function of the switching power
supply or the correct electric power supply (wall outlet) the unit may only
be used with internal electric power supply (battery).
Do not use the SIROLaser Xtend if a visual inspection shows that it has
been damaged.
Sirona Dental Systems GmbH
Operating Instructions SIROLaser Xtend
63 18 351 D3576
D3576.201.01.05.02    11.2018