2 Technical description

2
Technical description
Model designation
Nominal voltage
Nominal output of PerioScan
Water requirements
Operating conditions
Transport and storage conditions
Protection class
Degree of protection against electrical shock
Degree of protection against ingress of water and
foreign substances
Operating mode
Dimensions
Weight (without ultrasonic handpiece)
Year of manufacture
Tests/approvals
8
/158°F
/40°F
PerioScan
230 VAC, 50/60 Hz
40 VA
The rinsing/cooling water must be clean according to the
applicable regulations. It must not be contaminated.
Input pressure 1.5 – 5 bar
Dilution hardness 8° dH (1.43 mmol)
A water softener must be installed for water hardnesses
equal to or exceeding 12° dH (= 2.15 mmol).
Ambient temperature: 59 °F – 104 °F (15 °C – 40 °C)
Relative humidity: 30 % – 95 %
Air pressure: 700 hPa – 1060 hPa
Temperature: -40 °F – 158 °F (-40 °C +70 °C)
Relative humidity: 10 % – 95 %
Air pressure: 500hPa – 1060hPa
After a severe change in temperature, allow sufficient
time for acclimatization.
PerioScan: Protection class I
Type B applied parts
IP 20 – PerioScan is protected against ingress of solid
foreign substances > 12.5 mm Ø. Not protected against
ingress of water.
IP X1 – The foot switch is protected against dripping
water.
Continuous operation
Width: 150 mm, depth: 305 mm, height: 220 mm
Approx. 4 kg, with rinsing media tanks full
20XX
(on bottom side of unit)
This unit is intended exclusively for operating an ultra-
sonic handpiece approved by Sirona. Namely:
PerioSonic.
Sirona Dental Systems GmbH
Operating Instructions PerioScan
61 29 626 D 3496
D 3496.201.02.05.02