6 Indications, contraindications and medical precautions
6.2 List of preset indications
Application
Power
[W]
Desensitiza-
1,0
tion
Herpes
2,0
Periodontology
Periimplantitis
1,5
(Germ Reduc-
tion)
Perio. Germ
1,5
Reduction
Sulcular
2,0
Debridement
Surgery
Abscess
3,0
Epulis
3,0
Fibroma
3,0
Fistula
2,0
58
Modus
Duty
Fre-
Cycle
quenzy
[%]
[Hz]
CW
PF
50
10
PF
50
12
PF
50
10
PF
33
10
CW
CW
CW
PF
33
15
Operating Instructions SIROLaser Xtend
Fiber
Help Menu
[µm]
320
Apply tin fluoride solution as described in
scientific diode-laser studies on the sensitive
tooth areas, apply laser 2-4 mm away from
these regions - semicontact, total time per
area: 60 seconds.
Caution: Avoid contact to dentine, keep laser
tip always in motion!
320
Anesthetics not needed! Apply laser 1-3 mm
away from lesion for a few seconds - semicon-
tact, wave the laser fiber over the entire lesion.
Interrupt treatment briefly, if pain sensations
occurs.
320
Move the fiber tip around the implant gently up
and down with a sinuous movement, covering
the wall of the tissue.
Caution: Keep the laser tip always in motion!
320
Irradiate the whole pocket starting from the
deepest position using a meandering course to
cover all contaminated regions. Reduce
power, if pain sensations appear.
320
Move the fiber tip around the tooth gently up
and down with a sinuous movement, covering
the wall of the tissue. Reduce power, if pain
sensations appear.
Caution: Keep the laser tip always in motion!
320
Point directly with fiber towards the tissue
where the drainage canal is planned.
Start laser and perform the drainage.
Eject infected material. Perform laser treat-
ment again to remove residual infected mate-
rial.
Caution: Avoid bone contact during treatment!
320
Stretch the tissue and use laser tip like an
scalpel to excise the tissue.
320
Stretch the tissue and use laser tip like an
scalpel to excise the tissue. Depending on the
size of the fibroma energy can be adjusted un-
til desired cutting is achieved.
320
Carefully insert the fiber into the fistula canal,
start laser and retract fiber slowly from the
canal (1 mm/s). At bone contact retract the
fiber 1-2 mm before laser is activated.
Sirona Dental Systems GmbH
63 18 351 D3576
D3576.201.01.05.02    11.2018