5 Operation
5.1 Start the device for the first time
34
NOTE
Alphabetic and numeric letters, limitation to data input:
• Numbers are displayed with English decimals '.' for all languages/
regions.
• The power is displayed with one decimal place. Unit is watt (W).
• The time can be set-up as continuous or 1 to 9999 seconds. If
continuous is selected it counts up to maximum 9999 seconds after
activating application. If time is fixed it counts down. Unit is seconds
(s). If 9999 is exceeded treatment will stop and display jumps back to
treatment screen.
• The duty cycle is fixed on 50% or 100%. 100% is displayed as CW.
After the upgrade the duty cycle is adjustable in 1% steps, from 1% to
100%.
• The frequency can be entered by typing in the numbers or by
moving higher or lower by using 'plus' or 'minus'. If using 'plus' or
'minus' the frequency will be set from 1 Hz -10 Hz in increments of
1 Hz, from 10 Hz -100 Hz in increments of 10 Hz, from 100 Hz -1 kHz
in increments of 100 Hz and after the upgrade from 1 kHz -10 kHz in
increments of 1 kHz. If 0 Hz is set, CW is displayed. Unit is hertz (Hz).
• After upgrade: Between 1 kHz -10 kHz the duty cycle cannot be
below 10% and above 90% due to physical limits. So, at 1 kHz
-10 kHz any input below 10% will automatically be displayed and
used as 10% for the treatment as well as any input above 90% will
automatically turn into 90%.
• The energy will automatically be calculated and displayed with no
decimal place. Unit is joule (J). Before starting the application the to
be attained total energy will calculated and displayed if time is set. As
soon as application is started and time running the energy will be
accumulated as long as the laser is active.
• The average power wil automatically be calculated and displayed
with one decimal place. Unit is watt (W).
Clear screen before entering new parameters or data. Existing entries
will not be overwritten.
Newly generated applications or changed parameters of preset
applications will appear in red.