5 Operation
5.5 Main home screen
There occurred an error
about the EEPROM com-
munication which could
not be handled. Please
contact Sirona Dental Sys-
tems GmbH, your local
dental depot or your au-
thorized service center.
Confirmation via 'OK'.
The fiber could not be de-
tected. See chapter "Trou-
bleshooting ..."  [→ 65].
Confirmation via 'OK'.
52
Cooling fan is not working.
To avoid damage please
switch off the SIROLaser
Xtend and let it cool down
for a while. See chapter
"Troubleshooting ..." 
[→ 65].
There occured an error
because the user wants to
set the duty cycle below
10% or above 90% at a
frequency above 1 kHz.
Due to physical limits, this
is not possible. Confirma-
tion via 'OK'.
Operating Instructions SIROLaser Xtend
There occurred an error
due to overheating which
could not be handled.
Please contact Sirona
Dental Systems GmbH,
your local dental depot or
your authorized service
center. Confirmation via
'OK'.
There occured an error
because the user wants to
set the average power be-
low 0.05 W in the chopped
mode. Due to physical lim-
its, this is not possible.
Confirmation via 'OK'.
Sirona Dental Systems GmbH
There occurred an error to
the photo diode inside the
handpiece which could not
be handled. Please con-
tact Sirona Dental Sys-
tems GmbH, your local
dental depot or your au-
thorized service center.
Confirmation via 'OK'.
There occured an error
because the user wants to
set the average power be-
low 0.3 W in the peak-
pulse mode. Due to physi-
cal limits, this is not possi-
ble. Confirmation via 'OK'.
63 18 351 D3576