3.3.3 Turn the lung ventilator OFF
Stand-by mode. Hold the ON-OFF key a few seconds to turn the lung ventilator OFF.
3-20
During the switching OFF the system performs a clean of the oxygen inside
the lung ventilator which grants a 21% concentration on the inspiratory line.
This procedure grants a longer O
phase when the device is turned-on. The duration of this procedure
depends by the FiO2 present in the inspiratory line and can be of 60
seconds max.
Stand-by mode
In Stand-by mode the User
can set and/or edit all lung
ventilator parameters and
alarms limit relative to the
operating mode that you will
use on the patient that you
want to treat.
The system will ask you if you
want to Turn-Off the lung
ventilator (switched off).
Press NO: cancel the
command (the lung ventilator
returns to Stand-by mode).
Press YES: the lung ventilator
will be switched OFF.
sensor life and a quicker " Self Test "
2
User manual, DU3104101