Calibration Of Zero Point - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Multi-parameter controller dialog daca
Table of Contents

Advertisement

Calibration
Permitted calibration range
The permitted calibration range is
20 ... 300% of the sensor's rated
value.
Example of a shallow slope: Blocking
of the sensor membrane leads to a
low slope (low slope = low sensor
sensitivity)
Example of a steep slope: Surfactants
make the sensor membrane more
permeable, leading to a steeper slope
(steep slope = high sensor sensitivity)
9.4.3

Calibration of zero point

Necessity for calibrating the zero
point
Calibration of the zero point is not
generally necessary. A calibration of
the zero point is only necessary if the
sensor is operated at the lower limit of
the measuring range or if the 0.5 ppm
sensor version is used.
88
CAUTION!
Correct sensor operation / Run-in
period
Damage to the product or its sur‐
roundings
Correct measuring and metering
is only possible if the sensor is
working perfectly
Please read the operating manual
for the sensor
Please also read the operating
manuals for the fittings and other
components used
It is imperative that the run in
periods of the sensors are
adhered to
The run in periods should be
allowed for when planning com‐
missioning
It may take a whole working day
to run-in the sensor

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents