Selection Of The Ph Calibration Process - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Multi-parameter controller
Table of Contents

Advertisement

Selection of the pH calibra‐
9.1.1
tion process
To calibrate the controller there are three
available calibration processes:
2-point
n
Samples (1-point)
n
Data input
n
Calibration process selection
1.
Continuous display ➨
ð The calibration menu is dis‐
played
2.
Press
CAL pH
Calibration process
Buffer detection
Temperature
Fig. 31: Calibration process selection
ð The menu for selecting the cali‐
bration process appears.
3.
Using the arrow keys select the
desired menu entry and press the
key
ð The entry window appears.
4.
Using the arrow keys select the cal‐
ibration process and press the
key
5.
Then press
ð You can now start the selected
calibration process.
Samples
25.0 °C
A1025
Calibrating the controller
9.1.1.1
Setting the temperature
1.
Continuous display ➨
ð The calibration menu is dis‐
played
2.
Press
CAL pH
Calibration process
Buffer detection
Temperature
Fig. 32: Setting the temperature
ð The menu for selecting the
temperature for the calibration
process appears.
3.
Then press
ð You can now start the selected
calibration process.
Samples
25.0 °C
A1025
63