Calibration Of All Amperometric Measured Variables - ProMinent DULCOMETER D1Cb Series Instructions For Assembly And Use

Hide thumbs Also See for DULCOMETER D1Cb Series:
Table of Contents

Advertisement

Measured Variables and Operating Menus for Amperometric Sensors

6.5 Calibration of All Amperometric Measured Variables

WARNING!
Danger of incorrect metering
This can result in hazardous incorrect metering
During initial commissioning, the measured variable and the measuring range of the sensor must
be set prior to calibration. Refer to
Measuring Range " on page 37
calibration
zero p.: 4.00 mA
slope
6.50 mA/ppm
Fig. 29: Calibration of All Amperometric Measured Variables
Error message
Calibration not possible!
Gradient too low
Calibration not possible!
Gradient too high
DPD value too low
DPD > x.xx ppm
Calibration not possible!
* Please also note the operation manual for the respective sensor
56
Ä Chapter 4.1.2 "Selection of the Measured Variable and
calibration
DPD-value
1.55 ppm
calibration
Zeropoint
4.00 mA
Condition
Gradient too low
(< 20 % of standard gradient)
Gradient too high
(> 300 % of standard gradient)
DPD < 2 % of measuring range Repeat calibration after addition
< 3 mA
calibration
DPD-value
01.55 ppm
14.00 mA
calibration
zero p.:
tske over value?
4.00 mA
Remarks *
Repeat calibration
Repeat calibration
of metering medium or fit sensor
suitable for the process
Check sensor/cable
calibration
zero p.: 4.00 mA
slope
6.50 mA/ppm
A0018_GB