2221A User Manual
j.
CHECK — Timing accuracy is within 3% (0.24 division at the 10th
vertical graticule line), and linearity is within 7.5% (0.15 division over
any two of the center eight divisions). Exclude any portion of the
sweep past the 100th magnified division.
k.
Repeat parts i and j for the remaining SEC/DiV and time-mark
generator setting combinations shown in Table A-11 under the X I0
Magnified column.
Table A-11: Settings for Timing Accuracy Checks
SEC/DIV Switch
Setting
0.05
M S
0.1
M S
0.2
ps
0.5
M S
1 ps
2
ps
5
ps
10
ps
20
M s
50
ps
0.1
ms
0.2
ms
0.5
ms
1
ms
2
ms
5
ms
10
ms
20
ms
50
ms
0.1
s
0.2
s
0.5
s
Appendix C: Performance Verification
Time-Mark Generator Setting
Normal (XI)
X I0 Magnified
50
ns
M s
0.1
0.2
M S
0.5
M S
1 M S
2
M S
5
M S
10
M S
20
M S
50
M S
0.1
ms
0.2
ms
0.5
ms
1 ms
2
ms
5
ms
10
ms
ms
20
50
ms
0.1
s
0.2
s
0.5
s
10
ns
10
ns
20
ns
50
ns
0.1
M S
0.2
M S
0.5
M S
1
M S
2
M S
5
M S
M S
10
20
M S
50
M S
0.1
ms
0.2
ms
0.5
ms
1
ms
2
ms
ms
5
10
ms
20
ms
50
ms