Leica TCS SP8 SMD User Manual page 220

For fcs, flim and flcs
Table of Contents

Advertisement

Changing the Filter Cube
Because the locking screws are accessible from the outside, the filter cubes can be
replaced without removing the housing.
1. Abort the scanning operation.
2. Turn the key switch on the supply unit into the "OFF" position.
3. Switch off the APDs.
4. Detach the screws:
Figure 188:
5. Replace the filter cubes.
6. Retighten the locking screws.
7. Switch the APDs back on. It takes about 15 seconds until these are ready to operate. Do
not carry out any scanning operation during that time.
Correct orientation and exact positioning of the filter cubes are assured by a locating pin
arrangement of the retainer.
220
Adaptation at X1 port with SMD filter cube