Connecting Drain Tubing - Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual

Variable wavelength detectors
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
VWD-3100 and VWD-3400RS Variable Wavelength Detectors

4.3 Connecting Drain Tubing

To discharge liquid leaks and waste, the detector has a drain port at the bottom right of the
instrument.
Drain port
Fig. 15: Drain port
Direct liquid leaks to waste through the drain system of the UltiMate 3000 system. Direct
liquid leaks to waste via the drain system of the UltiMate 3000 system, using the
components from the drain kit. The kit is shipped with the UltiMate 3000 pumps and can
be ordered separately (part no. 6040.0005). The kit includes all required components and
detailed installation instructions. If there is more than one detector in your system and you
need an additional tee piece, you can find one in the accessories kit of the fluorescence,
multiple wavelength, or diode array detector.