4.3 Operating conditions

Temperature range: Operating
Relative humidity
Maximum operating altitude

4.4 Noise data

Note:
Data measured at ultimate vacuum 1 metre from end of the pump.
Table 4 Noise data
Pump Variant
E2M40 3 ph 400 V 50 Hz
E2M80 3 ph 400 V 50 Hz
E2M175 3 ph 400 V 60 Hz
E2M275 3 ph 400 V 60 Hz

4.5 Vibration data

Refer to pump instruction manual.

4.6 Electrical data

4.6.1 Electrical data: three-phase pumps

When you start a cold pump, the motor will draw the start‑up current shown in
Electrical data (three
reduce quickly as the motor reaches rated rotational speed. Within 5 minutes, as the oil and
pump warms up, the current drawn will slowly reduce to the full load current specified in
Table : Electrical data (three
When you start a warm pump, the motor will draw the start‑up current shown in
Electrical data (three
then immediately fall to the full load current specified in
pumps)
Electrical short‑circuit and ground‑fault protection of the pump will be provided by fitting
Class CC fuses of the values shown in
at the point of connection to the supply. If these are not available in your country of use,
Type aM European fuses of the same rating can also be used.
Table 5 Electrical data (three‑phase pumps)
Pump
E2M40
E2M80
E2M175
A36418880_A - Technical data
‑ phase pumps)
‑ phase pumps)
‑ phase pumps)
on page 11.
Nominal
Frequency
supply
(Hz)
(V)
400
50/60
400
50/60
400
50/60
+12 °C to +40 °C (53 °F to 104 °F)
on page 11 for several seconds. The current will then
on page 11.
on page 11 for up to 0.5 seconds. The current drawn will
Table : Electrical data (three
Table : Electrical data (three
Full load
Power
current
(kW)
1.1
2.2
5.5
Page 11
10% to 90%
1000 m
dB (A)
72
70
75
75
Table :
Table :
‑ phase pumps)
on page 11
Start‑up
current
(A)
(A)
2.55
11
4.8
14
11
52
‑ phase
Recommended
fuse rating
(A)
5
10
20

Advertisement