Maintenance: Pump - Agilent Technologies G8610A Operation Manual

Helium leak detector
Table of Contents

Advertisement

7.4.5 Maintenance: Pump

The Pump Menu screen displays the Fore pump type (Dry or Oil) - Factory configured.
Operating hours and Last Service are also displayed.
This menu also displays the installation of a diaphragm exhaust pump and its' operating
hours and Last Service Date.
Buttons and Fields
Last Serviced
Enable Exhaust Pump
screen
124
Agilent Technologies Helium Leak Detector
Figure 7-11 Maintenance: Fore Pump
Select the date on which the oil is changed or the tip seal is
replaced. The system then tracks the operating hours since the
pump was last serviced or when a new fore pump is installed. Lists
the Last Service date for the exhaust Pump and operating hours
This screen appears for dry pump models only. The service date is
changeable. Exhaust Pump only for models G8610B, G8611C,
G8612C, G8611B and G8612B.