Agilent Technologies G8610A Operation Manual page 21

Helium leak detector
Table of Contents

Advertisement

21
Agilent Technologies Helium Leak Detector
CAN/CSA-C22.2 No. 61010-1-12 - Safety Requirements for
WARNING
Electrical Equipment for Measurement, Control, and Laboratory
Use - Part 1: General Requirements
UL Std. No. 61010-1, 3rd Edition - Safety Requirements for
Electrical Equipment for Measurement, Control, and Laboratory
Use - Part 1: General Requirements
EMC Emissions and Immunity - ETSI EN 301 489-1 v2.2.0
(2017-03) Electromagnetic Compatibility and Radio Spectrum
Matters (ERM); Electromagnetic Comparability (EMC) standard
for radio equipment and services; Part1: Common technical
requirements
Any modifications on the part of the user are liable to cause
non-compliance with regulations or affect the EMC
performance and the safety of the product. Agilent cannot
be held responsible for consequences resulting from such
intervention.
Before powering the unit the first time, verify that the unit is
configured to operate for the local mains supply voltage.
The equipment can be damaged by:
All electrical connections must be performed by a qualified
electrician and must comply with national and local codes.
Opening the enclosure may expose hazardous voltages.
Always disconnect the power cord and interface cables
before opening the enclosure. Do not touch the power
inlet's contacts for at least 10 seconds after disconnecting
the power cord.
Electrical installation must include the appropriate branch
circuit (20 A maximum), with a long time delay and a
reliable earth ground.
Use only the power cord that was provided with your leak
detector. The use of extension cords is not recommended
and could result in damage to the equipment and loss of
warranty.
To avoid electric shock, connect the product power cord to
a grounded power receptacle. A power cord with a
grounding conductor is required.
Incorrect mains AC supply voltages
Radio Frequency (RF) and Electrostatic Discharge (ESD)
energy inputs that exceed the maximum ratings
Operating in very high temperatures or without adequate
ventilation
Immersion in liquids
Physical abuse