5.1.3.6 Spray Method

The spray method is for leak testing vacuum systems of all sizes. When properly connected
to the vacuum system under test, the Helium Leak Detector will locate and indicate the
relative size of leaks. Settings available in this mode include:
Entering set point values for audible and visual alarms.
Connecting the wireless remote - allows testing of large systems by one operator.
To access:
1. Select Test Configurations > Applications Setup.
2. Select Spray Method and press Set-Up and Figure 5-9 appears.
At this point Auto Zero < 0 is enabled.
Buttons and Fields
Set Point 1
Audio Set Point
Remote Control
Agilent Technologies Helium Leak Detector
Figure 5-9 Spray Method
Desired Alarmed Leak Rate value (> <) - This value can be adjusted
-10
from 10
to 10
ally programmed for greater than the acceptable leak rate limits.
Desired Audio Alarmed Leak Rate value (> <) - This value can be
adjusted from 10
own distinct tone if different from Set Point 1.
The wireless handheld remote can be activated and configured
(channel selection) within this application.
-4
. Set Point values in Spray applications are usu-
-8
-4
to 10
. The Audio Set Point triggers with its'