Section 4. Calibration Screen
4.1 Calibration Set-Up
❑
Press Calibration Menu screen.
The Calibration screen appears (Figure 4-1).
Use this screen to:
❑
❑
❑
❑
The calibration routine can only be executed while in Test/FineTest.
Choose the
Calibration
Type.
Select leak used
for autocal.
Agilent Technologies Helium Leak Detector
View the value and expiration date of the Internal Calibrated Leak
Enter the External Calibrated leak value
Select whether to use an Internal or external calibrated leak
Set the calibration mode: Full or Fast Calibration
Figure 4-1 Calibration Set-Up
The internal leak is a factory supplied helium leak internal to
NOTE
the leak detector. In order to perform an external calibration,
you must supply a separate calibrated leak and mount it on top
of the leak detector.
Shows the
temperature
compensated value
of the internal leak.
Enter the
External Leak
Value.
65