Agilent Technologies G8610A Operation Manual page 88

Helium leak detector
Table of Contents

Advertisement

5.1.3.8.2.2 Test Time Setup
1. Program the Test Time for some value that allows for correct characterization of the
necessary test time to reach at least one decade less than that of Setpoint 1 (reject level).
2. Input the desired Test Time.
5.1.3.8.2.3 Set Point 1 Setup
1. Input the Leak Rate value that is a failed/reject condition.
2. Typically, a greater than condition is programmed.
5.1.3.8.2.4 Set Point 2 Setup
You can choose between the two setpoints.
1. Input the Leak Rate value that is a failed/reject condition.
2. Typically, a greater than condition is programmed.
5.1.3.8.2.5 Using Set Point 1 and 2 together (Optional)
Some applications utilize a reference leak during the test cycle to validate that the leak
detector is still calibrated and sensitive to helium.
Set Point 1 is set up for the leak rate reject level (>).
Set Point 2 is set for the reference leak (<).
5.1.3.8.2.6 Pressure Set point (High & Low Limit) - Optional
High Limit - Input the high-pressure limit that must not be exceeded during the test.
Low Limit - Input the low-pressure limit that must be maintained throughout the test.
Files: The operator can store up to three setup configurations:
88
Agilent Technologies Helium Leak Detector
A Typical Auto Sequencer application uses either Set Point 1 or
NOTE
Setpoint 2 for the desired leak rate reject set point.
If the leak rate falls below the reference leak value (Set Point 2)
NOTE
at the end of the test cycle, the leak detector reports
Rejected/Hold. This type of a reject is a system level failure.
This could be related to a fixturing issue, operator error or
system performance.
AutoS 1-3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G8610bG8611aG8611cG8612aG8611bG8612b ... Show all

Table of Contents