Gross Test Screen - Agilent Technologies G8610A Operation Manual

Helium leak detector
Table of Contents

Advertisement

74
Agilent Technologies Helium Leak Detector
Figure 5-3 Gross Test Screen
Gross Test is activated in the Test Configurations > Advanced
Features menu.
Once activated, the GROSS TEST button appears on the
Home screen. Press GROSS TEST to enter the screen.
There is an audio tone for Gross Leak testing, which
increases as the leak increases.
Gross Test only works when the test port pressure is between
10 and 150 Torr and remains constant (± 2% between
previous and current pressure readings).
The Gross Leak test stops if the system goes into Backing,
Test or Fine TEST mode.
Activating the Auto Sequencer button terminates the Gross
Leak test.
The HOLD button is inactive while in Gross Leak test mode,
as the system is in roughing mode.