Agilent Technologies G8610A Operation Manual page 84

Helium leak detector
Table of Contents

Advertisement

5.1.3.8 Auto Sequencer
The Agilent Auto Sequencer facilitates programming a repetitive test cycle for leak testing
parts that have been sealed with helium inside. This can improve productivity and reduce
testing errors. Settings available in this mode include:
Program timers for the roughing and test modes.
Entering set point values for audible and visual alarms.
Establishing a lower level verification leak to improve reliability of the test.
Setting a pressure set point that saves time by rejecting a part if the pump down
characteristics fall outside of the norm.
The auto sequencer is used to automatically perform testing on a part, to maximize testing
throughput and to reduce the chances of accidentally accepting a bad part. The auto
sequencer is used in conjunction with the reject and audio set point settings, which provide
the criteria for pass or fail conditions. Within the auto sequencer feature, there are two
variables that must be set, rough time and test time. The ROUGH TIME setting indicates to
the system the maximum time it is allowed to be in roughing before the test fails. Ideally, the
system goes into the roughing phase and then proceeds to TEST. The TEST TIME setting tells
the system when to perform the test against the set point values. For example, you may set
20 seconds for the TEST TIME. At the end of 20 seconds, the system determines if the part
has passed or not and then makes an indication by either going to VENT (Pass) or going to
HOLD (Fail) and displaying the reject set point failure on the front panel.
Press Auto Sequencer Set-Up on the Menus screen.
The Auto Sequencer Set-Up screen appears (Figure 5-11). The leak detector can be
configured to automatically run through a complete test cycle
(Start/Rough/Test/Vent).
During the test cycle, the system ensures that during the Rough Time the
user-entered pressure transfer set points are met.
A Pass condition is indicated by the system reverting to the Vent mode upon
completion of the test cycle. A Fail condition is indicated by the system reverting to
the Hold mode, and all operating buttons being disabled except Vent. A failed
condition also results in a red activated set point box in the Set Point window
appearing on the Home screen, along with a reject set point activation upon
completion of the test cycle. The status of the reject set points are available through
the RS-232 line (see Appendix A "Communications Protocol").
On completion of an automated test cycle, the leak detector captures the leak rate
signal measured just prior to the end of the automated test cycle. This leak rate
appears on the Home screen until the next test cycle is initiated.
84
Agilent Technologies Helium Leak Detector

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G8610bG8611aG8611cG8612aG8611bG8612b ... Show all

Table of Contents