7.4.6.4 System Information

Press System Information on the Menus screen.
The System Information screen appears (Figure 7-14).
This screen gives details of the installed software version, spectrometer parameters
and the turbo frequency.
Condition
Ion Source Filament
See Section 7.4.3 "Manual Tuning"
on page 118.
Gain
Software Version and 
Agilent Web site
Agilent Technologies Helium Leak Detector
Figure 7-14 System Information
Table 7-2 System Information Screen Conditions
Filament 1 Off
Filament 1 On
Filament 2 Off
Filament 2 On
Gain (Gain Value)
Display
Filament 1 is selected and not lit.
Filament 1 is selected and lit.
Filament 2 is selected and not lit.
Filament 2 is selected and lit.
Displays the gain that applied to the
leak rate. The actual helium signal is
multiplied by the gain value to read
the correct, calibrated leak rate.
Shows the software version of the CPU
board.
Description