Recommended Additional Services - Agilent Technologies G8610A Operation Manual

Helium leak detector
Table of Contents

Advertisement

Figure 1-15 is a drawing of a Bench Mount Leak Detector with External IDP-15 Dry Scroll
Pump, Model G8612C with individual components shown.
Figure 1-15 Bench Mount Leak Detector with External IDP-15 Dry Scroll Pump, Model
1.3.1 Helium
Welding grade helium in a standard cylinder with pressure regulating valve and hose is
required for testing products and leak checking.

1.3.2 Recommended Additional Services

The following additional services are recommended, based on the application:
Connect an exhaust hose to the foreline pump and vent it outside of the room, or to
the facility's exhaust, to reduce the particulate count and the helium background
signal.
When using oil lubricated mechanical pumps, use an oil mist eliminator on the
exhaust port of the mechanical pump to reduce the oil mist from the exhaust.
Oil mist eliminators can, however, become saturated with oil causing slower
pumping speeds and higher helium backgrounds in the leak detector. The
replacement time interval of the oil mist eliminator is determined by whether the
user frequently operates at elevated roughing pressures, pumps on large volumes, or
frequently cycles the system. Refer to Table 8-4 on page 136 and Table 8-6 on
page 138 for the mist eliminator and replacement cartridge part number.
Dry nitrogen vent to keep the test port dry. Refer to Section 8.2 on page 135 for
more information.
Adequate circulation to prevent an increased helium background level.
36
Agilent Technologies Helium Leak Detector
G8612C

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G8610bG8611aG8611cG8612aG8611bG8612b ... Show all

Table of Contents