Appendix C. Specifications - Agilent Technologies G8610A Operation Manual

Helium leak detector
Table of Contents

Advertisement

Appendix C. Specifications

C.1 Specifications
The specifications for the leak detector are provided in Table C-1.
Specifications
Pumps
Fore Pump Type
Pump Speed,
l/min, 60/50 Hz
Minimum Detectable Leak,
atm-cc/sec
Internal Pump Version
Turbo Molecular Pump
Leak Units
Peak Pumping Speed
Helium Response Time
Test Port
Recommended Operating Ambient Conditions
Installation
Operating Temperature
Humidity
Heat Load
Air Pressure
Maximum system
vibration
Agilent Technologies Helium Leak Detector
Table C-1 Leak Detector Series Specifications
Agilent Technologies Helium Mass Spectrometer
Rotary Vane
DS40M
DS-302
36/30
285/237
Navigator™ Turbo molecular pump specifically designed for leak detection
atm-cc/sec, mbar-l/sec, torr-l/sec, Pa-m
Contra w/ DS40M – 0.5 l/sec for N
Contra w/ DS302 – 3.0 liters/sec for N
Contra w/ TS620 – 6.0 liters/sec for N
Contra w/ IDP-3 - 1.1 l/sec (60 Hz), 1.0 l/sec (50 Hz)
Midstage – 1.8 liters/sec for He at 50 or 60 Hz.
0.5 second, within 66% of peak reading within the given time. Per AVS 2.1.
NW-25; A 1 1 / 8 " (28 mm) ID Compression Port is available as an accessory.
Indoor use only. Altitude up to 2000 m (6,400'), Material Group III, POLLUTION DEGREE 2
according to EN61010-1 Standard, Class 1
System: +12° C to +45° C (wet and dry pumps); Fore pump limited on the low ends and turbo
pump limited on the high ends.
Maximum relative humidity (HR) 90%
1000 W (+/- 5%)
563 Torr (75 kPa) to 795 Torr (106 kPa)
3 G's (Turbo Limit)
Dry Scroll
TS-620
IDP-15
500/420
250/210
12
5 x 10
DS40M/IDP-3
3
3
/sec, std. ft
/year
and He at 60Hz in Contra (.42 l/sec @ 50Hz).
2
and He at 60Hz (2.5 l/sec @ 50 Hz).
2
and He at 60Hz (5.0 l/sec @ 50 Hz).
2
Versions
IDP-3
60/50
167

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G8610bG8611aG8611cG8612aG8611bG8612b ... Show all

Table of Contents