Table A-2 Non-Volatile Operating Parameters (Continued)
Command
INIT-4REJECT
INIT-AREJECT
HIGH-xREJECT
LOW-xREJECT
INIT-ROUGH
INIT-SYSLEAK
INIT-STDLEAK
INIT-TCTPZERO
INIT-TEMPFACTOR
INIT-TEST
LOCK-MS
148
Agilent Technologies Helium Leak Detector
Sets the test port pressure value for Reject Set Point #4. Preceded by a
two-digit leak rate mantissa with a decimal point after the first digit
followed by E-, and then by a two-digit exponent: the test port TC pressure
in millitorr.
Example 1.5E-02 INIT-4REJECT
Same as INIT-1REJECT, but for the Audio Reject Set Point.
Sets the reject set point to activate when the actual test value is higher than
the set point. Where x is a number from 1 to 4 or the letter A. For example,
the reject set point is set for 7.0E-05 and the actual test value is 9.2E-05.
Sets the reject set point to activate when the actual test value is lower than
the set point. Where x is a number from 1 to 4 or the letter A. For example,
the reject set point is set for 7.0E-05 and the actual test value is 4.3E-07.
Sets the Auto Sequencer rough time in seconds. Preceded by XX.
Sets the split flow leak value, system must be in TEST and the split flow
feature must be OFF when this parameter is set. Preceded by a two-digit
leak rate mantissa with a decimal point after the first digit followed by E-,
and then by a two-digit leak range exponent: the helium leak rate number
in atm cc/sec.
Example 3.4E-08 INIT-SYSLEAK.
Sets the value of the internal leak. Preceded by a two-digit leak rate
mantissa with a decimal point after the first digit followed by E-, and then
by a two-digit leak range exponent: the helium leak rate number in atm
cc/sec.
Example: 3.2E-08 INIT-STDLEAK.
Used to calibrate the low-pressure end of the test port thermocouple to a
known pressure higher than zero. Preceded by a number (XX) of mTorr
representing the test port pressure at the time of the command.
Sets the temperature correction factor that is printed on the internal cali-
brated leak label. Preceded by –10 to +10.
Example: -7 INIT-TEMPFACTOR
Sets the Auto Sequencer test time in seconds. Preceded by XX.
Enables the MS lock feature. If the system exceeds 150 mTorr, the system
goes from Fine test directly into Roughing mode (no Contra mode). If the
leak rate exceeds the 7 range, the over range symbol appears.
Parameter